Laboratory Bench Top Oven - AX

Close all | none



AX30

Max temp (°C) 250
Min temp (°C) Ambient +30
Temp uniformity (°C) ±5.0 @ 250°C
Heat-up time (mins) 23
Recovery time (mins) 3
Dimensions:
Internal H x W x D (mm)
295 x 300 x 320
Dimensions:
External H x W x D (mm)
440 x 590 x 465
Shelves fitted / accepted 2 / 4
Shelf loading each / total (kg) 10 / 20
Volume (litres) 28
Air changes / hr 65
Max power (W) 1000
Holding power (W) 342
Weight (kg) 24

AX60

Max temp (°C) 250
Min temp (°C) Ambient +30
Temp uniformity (°C) ±5.0 @ 250°C
Heat-up time (mins) 25
Recovery time (mins) 3
Dimensions:
Internal H x W x D (mm)
395 x 400 x 420
Dimensions:
External H x W x D (mm)
540 x 690 x 565
Shelves fitted / accepted 2 / 6
Shelf loading each / total (kg) 10 / 30
Volume (litres) 66
Air changes / hr 28
Max power (W) 1500
Holding power (W) 465
Weight (kg) 37

AX120

Max temp (°C) 250
Min temp (°C) Ambient +30
Temp uniformity (°C) ±5.0 @ 250
Heat-up time (mins) 26
Recovery time (mins) 3
Dimensions:
Internal H x W x D (mm)
495 x 500 x 520
Dimensions:
External H x W x D (mm)
640 x 790 x 665
Shelves fitted / accepted 2 / 8
Shelf loading each / total (kg) 10 / 40
Volume (litres) 128
Air changes / hr 14
Max power (W) 2000
Holding power (W) 622
Weight (kg) 55

Please note:
- Uniformity is measured in an empty chamber with vents closed, after a stabilisation period
- Shelf loadings are based on evenly distributed weight
- Maximum power and heat up times based on a 240V supply
- The uniform volume is smaller than the total chamber volume
- Minimum operating temperature is ambient +30°C

Content may be subject to modifications or corrections